The room temperature of the selected installation place should be 10℃ - 35℃. The relative humidity should be below 85% (45% - 65% recommended). The selected place should be protected from corrosive gas, strong magnetic field, direct sunlight and vibration, and also should be equipped with ventilation device.
The worktable selected should be leveled and stable, and capable of supporting the weight of analyzer and that of other accessories attached with the analyzer.
Dimensions of the worktable: length of not less than 1.5m, width of not less than 0.8m, height of 0.5m - 0.8m. The distance between the wall and the worktable should be over 0.5m, so as to facilitate the connections of gas soft tubes and cables.
A network router or switch may be needed for computer-GC communication or remote diagnosis.