EEU Specification for Pensky-Martens closed Cup Flash Point Tester
Instrument:
Semi-Automated Pensky-Martens closed Cup Flash Point Tester for testing Flash Point of Mineral Insulating oils as per ISO-2719 / ASTM D-93.
Minimum Requirements
Temperature Range: Up to 400 degree
Stirring speed: Adjustable 60 to 250 RPM
Testing Mode: Built in test methods - ISO-2719, ASTM D-93 (A & B) and User Custom setting.
Report: Printer with time Stamping
Ignition: Interchangeable Gas Ignition and Electrical Ignitor
Barometric Pressure: Automatic Barometric Display and Correction
Power Supply: 220 V Single Phase AC with 50 Hz frequency
Display: Real time on Screen display of Test Progress. LCD Display and Alphanumeric Keypad
Memory: Built in memory up to 50 Readings.
Cooling: Air forced
Heating Rate: Selectable. Minimum of 4 different heating rates, Including fast pre heat mode.
Certificate: Instrument shall be supplied along with the confirmatory certificate from manufacturer.
Standard Flash Point Liquid - NIST: NIST Traceable Reference Liquid with certificate.
Safety: Auto shut off with Error Reporting
Flash Detection: CRC thermocouple
Temperature Sensor: PT-100 in stainless steel with calibration certificate.
Special Conditions:
1) The Supplier shall supply NIST Traceable certified flash point reference material ( 3 No's x 80 ml each) having nominal values of 140 degree.
2) The vendor shall offer warranty of minimum one year.
BOQ
Semi Automated PMCC flash point tester for testing of Flash Point of Mineral Insulating oils as per ISO-2719 / ASTM D-93 and shall consist of minimum following item. The PMCC Flash Point Tester shall meet the Technical Specification enclosed herewith.
1) Sample Cup with Cover
2) Temperature Sensor
3) Igniter - Gas and Electric
4) Flash point Detection Sensor
1 Set
Spares:
1) Electrical Igniter
2) Gas Igniter
3) Sample Cup with Cover
4) Temp. Measurement Sensor
5) Flash Point Detection Sensor.
1 No Each
Annual Maintenance Contract
(After expiry of warranty): 5 Years