Specification for HPLC (High Performance Liquid Chromatography) based test set to carry out Furan Analysis of insulating oils as per IEC 61198-93 and also as per ASTM-5837-99.
6. OTHER REQUIREMENTS
a) Power requirements: All the system components shall be operated with 220V, 1ph, 50Hz, AC. Suitable power cards shall be supplied with the relevant kits.
b) System integration: It is the responsibility of the vendor to ensure integration of entire test system. He shall offer all the essential components to carry out analysis, compatible parts / products and mention clearly the mandatory parts required to match specifications / carry out testing as per ASTM-D-5837-1999
c) Compliance to EACH of the specifications mentioned above (including spares)shall be supported by the original brochures / communications from original manufacturer of the equipment. Mere endorsement on the quotation by the bidder shall not be accepted as sufficient evidence of compliance.
d) IQ/OQ/PQ tests shall be conducted as a part of commissioning and a certificate of compliance shall be issued as supporting documentation. Installation of system shall be deemed complete only after the establishment of methods and demonstration of actual sample testing as per ASTM-D-5837 / IEC-1198-1993.
e) Supplier shall conduct training on instrument as well as application to familiarize laboratory staff to the system.