EEU Specification for GC (Gas Chromatograph)
Reference Specification for test setup having GC (Gas Chromatograph) with HEAD SPACE AUTO SAMPLER to carry out Dissolved Gas Analysis of insulating oils as per International Standard ASTM-3612-02 Method - C
EEU only arranges the following materials
High Purity Gas cylinders - Argon, Hydrogen and Zero air
Calibration gas mixture with certificates of gases of interest of various concentrations.
Single Phase 230 V AC at 50 Hz frequency UPS power

Note:
Installation and commissioning of the test set up shall be deemed complete only after the establishment of methods, calibrating the instruments, establishing minimum detection limits, establishing repeatability & reproducibility, preparing multi point calibration tables and demonstration of actual sample testing as per ASTM-D-3612/2002, Method - C.