ASTM D5483 Test Method for Oxidation Induction Time of Lubricating Greases
ASTM D5483 Standard Test Method for Oxidation Induction Time of Lubricating Greases by Pressure Differential Scanning Calorimetry
8. Calibration
8.1 Sample Temperature Calibration:
8.1.1 Weigh approximately 10 mg of indium into an aluminum sample pan, insert a lid and crimp the lid to the pan using the encapsulation press. Place the crimped pan onto the sample platform in the pressure cell. Seal an empty pan in the same manner and place it on the reference platform. Set the cell cover in place and close the cell.
8.1.2 Open the oxygen cylinder valve slightly and set a pressure of 3.5 ± 0.2 MPa (500 ± 25 psig) on the cell inlet line with the pressure regulator. Partially open the inlet valve on the cell and allow the pressure to slowly build up in the cell. This should require approximately 2 min. Using the outlet valve, adjust the oxygen purge rate through the flowmeter to 100 ± 10 mL/min. The open position of these valves should remain fixed during the test.
8.1.3 Set the thermal analyzer to heat from ambient temperature (approximately 22°C) to 180°C) at a programmed rate of 10°C/min. After completion of the run, measure the melting temperature of the indium. If the melting temperature differs from 157.4 ± 0.2°C (see Note 4), correct the difference by using either the hardware or software calibration procedure described in the manufacturer's instruction manual. If the hardware calibration procedure is used, the temperature correction should be performed under 3.5 MPa (500 psig) oxygen pressure with a 100 mL/min purge rate. A typical melting calibration curve is shown in Fig. 2.
NOTE 4 - The melting temperature of indium is 156.6°C at atmospheric pressure, but has been found to be elevated to 157.4°C under the conditions of this test method, 3.5 MPa (500 psig) of oxygen.
8.2 Temperature Controller Calibration:
8.2.1 Remove both the sample pan and the reference pan from the cell, then close the cell. Slowly pressurize the cell with 3.5 ± 0.2 MPa (500 ± 25 psig) oxygen and adjust the purge rate to 100 ± 10 mL/min using the cell outlet valve. Select the desired test temperature (either 210, 180, or 155°C).
8.2.2 Program the cell to maintain the selected test temperature. If, after 10 min, the displayed cell temperature differs by more than ±0.2°C from the selected temperature, slowly adjust the temperature controller until they agree. After making an adjustment, wait at least 5 min to make certain that the temperature is stable before continuing.
8.3 Cell Base Pressure Gage Calibration:
8.3.1 The calibration should be conducted using a calibrated pressure transducer or a previously calibrated gage according to the pressure cell manufacturer's instructions.