ASTM D1533 for water in insulating liquids by Coulometric KF Titration
ASTM D1533 standard test method water in insulating liquids by Coulometric Karl Fischer Titration
4. Significance and Use
4.1 Electrical characteristics of an insulating liquid may be affected deleteriously by excessive water content. A high water content may make a dielectric liquid unsuitable for some electrical applications due to deterioration of properties such as the dielectric breakdown voltage.

4.2 These tests are suitable for use in acceptance specifications, in control of processing, and in evaluating the condition of dielectric liquids in service.

5. Interferences
5.1 Compounds such as aldehydes, ketones, free halogens, most acids, and oxidizing or reducing agents may interfere with coulometric Karl Fischer titrators. If a drifting end point is noted, an alternative solvent system or another titration method is warranted. If this drifting end point cannot be corrected, the water values should be regarded as suspect. A detailed discussion of interfering substances can be found in the treatise on aquametry.

5.2 Studies have shown that the water content of an insulating liquid sample may be influenced significantly by the sample container. A sample may either gain or lose water on storage in a glass container depending upon the initial water content of the sample, the manner in which the container is cleaned and dried, and the length of storage time before analysis. In addition, sample bottles should not be dried at temperature in excess of 110°C and should be rinsed with the liquid being tested prior to taking the test specimen.

5.3 Erroneous low readings may be obtained if previous spent oil test specimens are not removed from the solvent system on a routine basis. Excess oil may not mix thoroughly with the solvent system thus preventing the total water content of that test specimen from being measured properly. Stirring should be at such a rate that an oil layer will not form on top of the reagent. If such an oil layer does form while at the instruments maximum stirrer speed, stop testing and remove the oil layer. If accurate results can not be obtained, the solution should be discarded.

5.4 Upon setting up of the titration vessel and solvent system, the walls of the titration vessel should be wetted by swirling the solvent system solution around in the vessel.